Eutech CyberScan PCD 650 Multi-Parameter

Darmatek

Eutech CyberScan PCD 650 Multi-Parameter

Product Features

A Eutech Star Buy! Combining the strengths of all CyberScan 600 series meters, the CyberScan PCD650 is our top-of-the-line waterproof handheld that enables concurrent measurement of pH, conductivity, dissolved oxygen and temperature, all at the same time.
  • Simultaneously measures and displays four parameters
  • ±1% full scale accuracy at 3-decimal resolution across a wide measurement range
  • Ion-measurement capabilities with 3-digit resolution
  • Accepts two-cell and four-cell conductivity probe to measure up to 500 m/s — meter measures pure water!
  • Selectable multi-point calibration for more accurate conductivity measurements or single-point calibration for quick, convenient calibration with no mistakes
  • Measures 90.00mg/L in DO concentration and 600% in DO saturation—one of the widest dissolved oxygen measurement ranges offered in the handheld market today
  • Fast and easy DO calibration with 100% calibration option against atmospheric air or 0% calibration against a zero solution
  • Salinity, Temperature and Pressure Compensation—accurate readings under varying conditions
  • No warm up time required with galvanic system of measurement
  • One-Glance-Tells-All Screen Display with Backlight
  • Electrode status report, calibration data and simultaneous display of up to 4 parameters
  • Rugged and IP67-rated waterproof for applications in harsh environments. Meter floats on water!
  • Easy-to-understand prompt messages for easy set-up, calibration and trouble-shooting
  • Recorded readings individually stamped with time and date in GLP compliant format
  • Auto-logging meter saves readings automatically within seconds to prevent data loss
  • HI-LO limit alarm for easy quality control – Meter alerts when readings fall outside selected ranges
  • Wireless Information Technology: No wires, no cables—Transfer data from meter to computer with the press of a button 
  • Non-Volatile Memory protects information and meter settings, even when batteries run out
  • Four Levels of Password Protection prevents tampering or accidental data deletion
  • Calibration Alarm: No more out-dated calibration settings

Applications

Environmental: Test water quality, monitor health of aquatic ecosystems, survey surface and ground water drinking supplies, and meet EPA regulations.

Water Quality Testing: For analyzing water hard water, untreated water, drinking water, effluent water, pool water and incoming process water. Ideal for all types of quality assurance and water quality testing.

Industrial: For checking metal finishing, cooling tower water, printing fountain solutions, boiler water, brines, rinse tanks, ponds, pollution control, recirculating systems, waste water and industrial process systems. Useful for food analysis and quality assurance testing.
Aquaculture: Use to monitor oxygen levels in catfish and shrimp farming; game stocking ponds; ornamental fish tanks and ponds; and other fish farming applications.

Specifications

Product Specification
Description
pH Range
-2.000 to 20.000 pH
Resolution & Accuracy
0.1/0.01/0.001 pH & ±0.002 pH
No. of pH Calibration Points
1 (Offset) to 6 points 
pH Buffer Options
USA, NIST, DIN, PWB, Custom
mV Range
±2000.0 mV
Resolution & Accuracy
0.1 mV & ±0.2 mV
Ion Concentration Range
0.001 to 19900
Resolution & Accuracy
2 or 3 digits & ±0.5% FS (mono); 1% FS (div)
No. of Ion Calibration Pts
Up to 8
DO Range
0 to 90.00 mg/L or ppm / 0 to 600%
Resolution & Accuracy
0.01 mg/L or ppm / 0.1% ; ±0.2mg/L / ±0.2%
Salinity Correction
Automatic
Resolution & Accuracy
0.0 to 50.0ppt & 0.1 ppt
Barometric Pressure Correction
Automatic correction with built-in sensor
Resolution & Accuracy
450 to 825 mmHg / 59.9 to 109.9 kP &
0.1 mmHg or 0.1 kPa
Conductivity Range
.. to 500 mS/cm
Resolution & Accuracy
0.05% FS & ±1% FS + 1 LSD
Cell Constant
0.010 to 10.000
Temperature Coefficient
Linear & Pure
TDS Range
.. to 500.0 ppt
Resolution & Accuracy
0.05% FS & ±1% FS +1 LSD
Salinity Range
0 to 80 ppt
Resolution & Accuracy
0.05% FS & ±1% FS +1 LSD
Temperature Range
-10.0 to 110.0 °C / 14.0 to 230.0 °F (Meter)
Resolution & Accuracy
0.1 °C/°F & ±0.5°C/ ±0.9°F
Temperature Compensation
Automatic / Manual (0 to 100 °C)
Memory
500 Data Sets with Date/Time stamping &
Security password protection

Output
IrDA Wireless Communications
Display
Multi-data Back-lit dot matrix; 7.2 cm (H) X 5.4 cm (W)
Special Functions
Auto-Calibration; Data-logging options with time/date indexed,  GLP compliance; Simultaneous Display and measurement of 4 parameters; Cal-due alarm; pH slope/offset Display & Self-diagnostic
Power
4 X 1.5 V "AA" Alkaline batteries or 9 V DC adapter, 500 mA

AZ 8602 Combo PH/COND./SALT/D.O.

Darmatek

AZ 8602 Combo PH/COND./SALT/D.O.

Features :

Large LCD with multiple parameters measurement in one meter
Easily measure pH/Conductivity or salinity/Dissolved oxygen in  one meter by using different probes in turns
Handheld fashion housing with IP67 waterproof design
Auto temperature compensation
Auto ranging for conductivity measurement
Multiple points calibration: 3 points for pH, 1 point for D.O.,
4 points for conductivity.
Manually altitude & salinity compensation for D.O. measurement
99 points memories w/ recall function
Low cost replacement probe
Temperature unit °C/°F is switchable
Powered by 4pcs AAA batteries
1 hour auto power off

Application:
  • excellent to use in fresh water aquafarm
  • excellent to use in sea water aquafarm
Specifications

pH Range
2.00 ~ 12.00
pH Accuracy
± 0.1
pH Resolution
0.01
Cond. Range
0 ~ 199.9, 0 ~ 1999uS/cm;
0 ~ 19.99, 0 ~ 69.9mS/cm
Cond. Accuracy
± 1% F.S ± 1 digit
Cond Resolution
0.1uS/cm, 1uS/cm;
0.01mS/cm, 0.1mS/cm
Salinity Range
0 ~ 10.00ppt; 0 ~ 42.0ppt (SEA WATER)
Salinity Accuracy
± 1% F.S ± 1 digit
Salinity Resolution
0.01ppt, 0.1ppt
D.O. Range
0.0 ~ 199.9% (0.0 ~ 30.0mg/L)
D.O. Accuracy
± 3% of F.S. + 1 digit
D.O. Resolution
0.1
Temp. Range
0 ~ 60.0°C
Temp. Accuracy
± 0.5°C
Temp. Resolution
0.1
Compatible Probe
PH PROBE, COND. PROBE AND D.O. PROBE COULD BE USED ON SAME METER IN TURNS
LCD Size (mm)
32.5 (H) x 54 (W)
Operating Temp. & RH%
0 ~ 50°C, Humidity <80%
Storage Temp. & RH%
0 ~ 60°C, Humidity <90%
Dimension (mm)
169 (L) x 78.3 (W) x 43.4 (H)
Weight
200g
Power
4Pcs AAA Batteries
Standard Package
Meter/Manual/Carry Case/Battery
Standard Probe
PH Probe/Cond. Probe/D.O Probe/D.O. Electrolyte/2 Membrane Set/Cable Clip (User can decide which probe to be included as standard)

AZ 86505 pH/ORP/Cond./TDS/Salinity Meter

Darmatek

AZ Instrument 86505 pH/ORP/Cond./TDS/Salinity Meter
  • Super large LCD with multiple parameters display
  • Powered by adaptor for continuously using
  • Durable electrode holder for left & right side placement
  • Measuring stability indicator
  • Auto or manual temperature compensation
  • Auto ranging or manual ranging
  • Up to 5 points calibration points
  • Data hold/memory maximum/memory minimum function
  • Calibration & electrode information review
  • 99 points memories w/ real time stamp
  • PC output for data donwload
Specifications

pH Range
0.00 ~ 14.00
pH Accuracy
± 0.02
pH Resolution
0.01
mV Range
± 1999mV
mV Accuracy
± 0.2mV(-199.9 ~ 199.9mV) or ± 2mV(others)
mV Resolution
± 0.1mV(-199.9 ~ 199.9mV) or ± 1mV(others)
Cond. Range
0 ~ 19.99, 0 ~ 199.9, 0 ~ 1999uS/cm;
0 ~ 19.99, 0 ~ 199.9mS/cm
Cond. Accuracy
± 1% F.S ± 1digit
Cond. Resolution
0.05% of Full Scale
TDS Range
0 ~ 19.99, 0 ~ 199.9, 0 ~ 1999ppm;
0 ~ 19.99, 0 ~ 199.9ppt
TDS Accuracy
± 1% F.S ± 1digit
TDS Resolution
0.05% of Full Scale
Salinity Range
0 ~ 11.38ppt
0 ~ 80.0ppt(NaCI)
Salinity Accuracy
± 1% F.S ± 1digit
Salinity Resolution
0.1
TDS Factor
0.3 ~ 1.00
Temp. Coefficient
0 ~ 10.0%°C
Temp. Range
0 ~ 80.0°C
Temp. Accuracy
± 0.5°C
Temp. Resolution
0.1
Compatible Probe
PH
ORP
COND.
LCD Size (mm)
40(H) x 105(W)
Operating Temp. & RH%
0 ~ 50°C, Humidity < 80%
Storage Temp. & RH%
-20 ~ 60°C, Humidity < 90%
Dimension (mm)
260(L) x 168(W) x 58(H) mm
Weight
150g
Power
9V Adaptor
Standard Package
Meter/USA Plug Type Adaptor/Manual/Probe/Solution/Cable/Software CD/Probe Stand

Lutron WA-2015 pH/ORP, DO, CD/TDS METER, Bench type

Darmatek

LUTRON WA-2015 pH/ORP, DO, CD/TDS METER, Bench type

Description

Bench type, RS23/USB, Data logger
pH/ORP, DO, CD/TDS METER
Model : WA-2015
* Professional bench meter for multi-function: pH/ORP, Conductivity/TDS, Dissolved oxygen.
* pH : 0 to 14.00 pH.
* ORP (mV): 1,999 mV.
* Conductivity: 200 uS/2 mS/20 mS/200 mS.
* TDS: 200/2,000/20,000/200,000 PPM.
* Large LCD with green color backlight.
* ATC (automatic temperature compensation).
* Real time 16,000 no. data logger.
* Data hold, Record (Max., Min.), RS232
* Optional electrode : pH electrode, ORP electrode, CD/TDS probe, Dissolved oxygen probe, pH ATC probe.
* DC 1.5V (UM-3, AA) x 8 PCs or DC 9V adapter in.

HANNA HI 981504 pH/TDS/Temperature Monitor


Darmatek

HANNA HI 981504 pH/TDS/Temperature Monitor
  • Simultaneous display of pH, TDS and temperature
  • Continuous measurements
  • Measure in °C or °F
  • Three, backlit displays
Set-up is simple. Install the HI 981504 near the sample to be tested, plug the indicator in and immerse the probes. pH, TDS and temperature measurements will be simultaneously displayed on three backlit LCDs.
Users can easily select the temperature unit (°C or °F) on the back panel.

The HI 1286 gel-filled pH electrode is provided with a waterproof sleeve to protect the BNC connector. The unique design of the electrode provides longer life in aggressive solutions. The HI 7634 TDS probe is easy to clean and requires little maintenance. Measurements are accurate and the meter can be calibrated at one or two points for pH and at a single point for TDS. Temperature is factory calibrated.

Order Information:

HI 981504/5-01 (115V) and HI 981504/5-02 (230V) are supplied with HI 1286 pH electrode, HI 7634 TDS probe (fixed), temperature probe (fixed), HI 70004 pH 4.01 buffer solution sachet, HI 70007 pH 7.01 buffer solution sachet, HI 70032 1382 ppm calibration solution sachet, HI 700661 electrode cleaning solution sachet (2), screwdriver, 12 VDC adapter and instructions.
HI 981504/7-01 (115V) and HI 981504/7-02 (230V) are supplied with HI 1286 pH electrode, HI 7634 TDS probe (fixed), temperature probe (fixed), HI 70004 pH 4.01 buffer solution sachet, HI 70007 pH 7.01 buffer solution sachet, HI 70442 1500 ppm calibration solution sachet, HI 700661 electrode cleaning solution sachet (2), screwdriver, 12 VDC adapter and instructions.

Specifications

Range
pH
0.0 to 14.0 pH
  TDS
0 to 1990 ppm
  Temperature
-10.0 to 60.0°C / -14.0 to 140.0°F
Resolution
pH
0.1 pH
  TDS
10 ppm
  Temperature
0.1°C / 0.1°F
Accuracy
@ 20°C

pH
±0.2 pH
  TDS
±2%
  Temperature
±0.3°C / ±0.5°F
Calibration
pH
manual two points through trimmers
  TDS
manual one point through trimmer
TDS Factor
HI 981504/5
0.5
  HI 981504/7
0.7
Probes
pH
HI 1286 PEI body pH electrode with 2 m (6.6’) cable (included)
  TDS
HI 7634 TDS probe (fixed)
  Temperature
stainless steel with 2 m cable (fixed)
Temperature
Compensation


automatic, 5 to 50°C (41 to 122°F), for TDS readings only
Power Supply

12 VDC adapter (included)
Environment

0 to 50°C (32 to 122°F); 95% RH
Dimensions/Weight

160 x 110 x 35 mm (6.3 x 4.3 x 1.4”)/560 g (1.2 lbs.)

HANNA HI 98204 Water Test Portable pH/ORP/EC/°C Field Meter

Darmatek

HANNA HI 98204 Water Test Portable pH/ORP/EC/°C Field Meter
  • Four instruments in one
  • Lightweight with handle
  • Easy to take measurements
  • Long battery life
Water Test provides quick and reliable measurement of the four most important variables of water analysis: pH, ORP, conductivity and temperature.

The lightweight and compact body makes it perfect to use in the plant as well as in the field. Superior in design and construction, Water Test is made of rugged VALOX® material which offers excellent protection from typical environmental conditions.

Water Test is designed with such simplicity that even non-technical users can operate it: fill the base of Water Test with the sample to be measured, press "RANGE" to select the mode and then simply read your measurement!
The rugged construction and simple operation make Water Test ideal for  water treatment analysis, wastewater tests, ecological studies, aquaculture and hydroponic applications.

Order Information:
HI 98204 (Water Test) is supplied with measuring compartment, batteries and instructions.

Specifications

Range pH 0.0 to 14.0 pH
Range mV ±1000 mV
Range EC 0 to 1999 µS/cm
Range Temperature 0.0 to 60.0°C
Resolution pH 0.1 pH
Resolution mV 1 mV
Resolution EC 1 µS/cm
Resolution Temperature 0.1°C
Accuracy pH ±0.2 pH
Accuracy mV ±5 mV
Accuracy EC ±2% F.S.
Accuracy Temperature ±1°C
Calibration pH manual, 2 points
Calibration mV factory calibrated
Calibration EC manual, 1 point
Calibration Temperature factory calibrated
Battery Type / Life
3 x 1.5V / approx. 200 hours of continuous use
Environment
0 to 50°C (32 to 122°F); RH max 95%
Dimensions
190 x 85 x 85 mm (7.5 x 3.3 x 3.3 ")
Weight
260 g (9.2 oz.)

HANNA HI 9813-6 Portable pH/EC/TDS/Temperature Meter


Darmatek

HANNA HI 9813-6 Portable pH/EC/TDS/Temperature Meter
with CAL CHECK™ Feature
  • Water resistant
  • On-screen tutorial messages for calibration
  • Exclusive CAL CHECK™ feature; Alerts users of calibration status
  • BEPS (Battery Error Prevention System) alerts the user in the event that low battery power could adversely affect readings
  • Switch parameters at the touch of a button
  • Automatic Temperature Compensation for pH and conductivity
  • Simple calibration; Dials located on the front panel makes calibration quick and simple, even for non-technical users
  • Battery % displayed on startup
  • Fast response multiparameter probe
HI 9813-6 is a versatile, water resistant, multiparameter portable instrument specifically designed for agricultural applications such as hydroponics, greenhouses, farming and nurseries.

This meter features an large LCD that clearly displays the parameter being measured as well as calibration instructions. Calibration is fast and easy with knobs located on the front panel of the instrument.
HI 9813-6 includes all the features of the HI 9813-5 while incorporating our exclusive CAL CHECK™ feature. CAL CHECK™ allows the user to easily check the pH probe calibration status at any time.

This instrument utilizes the HI 1285 series pH/EC/TDS/temperature probe. This probe features a fiber junction and gel electrolyte making it ideal for fertilizer solutions.

Order Information:

HI 9813-6 is supplied with HI 1285-6 multiparameter probe, HI 70007 pH 7.01 calibration solution sachet, HI 70442 1500 ppm (mg/L) calibration solution sachet, HI 70031 1413 µS/cm calibration solution sachet, 700521 electrode cleaning solution sachets (2), battery, instructions and rugged carrying case.

Specifications

Range
pH
0.0 to 14.0 pH
  EC
0.00 to 4.00 mS/cm
  TDS
0 to 1999 ppm (mg/L)
  Temperature
0.0 to 60.0°C
Resolution
pH
0.1 pH
  EC
0.01 mS/cm
  TDS
1 ppm (mg/L)
  Temperature
0.1°C
Accuracy @ 20°C
pH
±0.1 pH
  EC
±2% F.S. mS/cm
  TDS
±2% F.S. ppm
  Temperature
±0.5°C
TDS Conversion Factor

0.56 to 0.78 ppm = 1 µS/cm (according to TDS 442 curve)
pH Calibration

manual, one point (all parameters except temperature)
Temperature Compensation

automatic 0 to 50°C (32 to 122°F) with ß=2%/°C
Probe

HI 1285-6 polypropylene body, pre-amplified multiparameter probe with CAL CHECK™ compatibilty, internal temperature sensor, 8-pin DIN connector and 1 m (3.3’) cable (included)
Battery Type / Life

9V / approximately 150 hours of continuous use
Environment

0 to 50°C (32 to 122°F); 100% RH
Dimensions / Weight

145 x 80 x 36 mm (5.7 x 3.1 x 1.4”) / 230 g (8.1 oz.)

HANNA HI 9811-5 Portable pH/EC/TDS/Temperature Meter

Darmatek

HANNA HI 9811-5 Portable pH/EC/TDS/Temperature Meter
low range EC & TDS
  • Water resistant
  • On-screen tutorial messages for calibration
  • Switch parameters at the touch of a button
  • Automatic Temperature Compensation for pH and conductivity
  • Simple calibration; Dials located on the front panel makes calibration quick and simple, even for non-technical users
  • BEPS (Battery Error Prevention System) alerts the user in the event that low battery power could adversely affect readings
  • Battery % displayed on startup
  • Fast response probe included
HI 9811-5 is a pH/EC/TDS waterproof meter designed for simplicity in taking pH, µS/cm, ppm (mg/L) and temperature measurements. both the HI 9811-5 and HI 9812-5 are ideal for hydroponics, greenhouses, farming and ground water applications.
Due to the built-in temperature sensor, conductivity readings are automatically compensated for temperature changes. The temperature coefficient is fixed at 2%/°C.
No probe changes are required when switching your measured parameter between pH, conductivity and TDS. These multiparameter meters reduce the number of instruments required for daily water quality analysis.

The specially engineered HI 1285-5 pH/EC/TDS and temperature probe utilizes a fiber junction and gel electrolyte which provide a fast response and reduced contamination. This combination makes the HI 1285-5 particularly designed to withstand harsh environments such as fertilizer solutions.

Order Information:

HI 9811-5 is supplied with HI 1285-5 multiparameter probe, HI 70007 pH 7.01 calibration solution sachet, HI 70032 1382 ppm (mg/L) calibration solution sachet, HI 70031 1413 µS/cm calibration solution sachet, HI 700661 electrode cleaning solution sachets (2), battery, instructions and rugged carrying case.

Specifications

Range
pH
0.0 to 14.0 pH
  EC
0 to 6000 µS/cm
  TDS
0 to 3000 mg/L (ppm)
  Temperature
0 to 60°C
Resolution
pH
0.1 pH
  EC
10 µS/cm
  TDS
10 mg/L
  Temperature
0°C
Accuracy
pH
±0.1 pH
  EC
±2% F.S.
  TDS
±2% F.S.
  Temperature
±0.5°C
TDS Conversion Factor

0.5 mg/L = 1 µS/cm
Calibration

manual at 1 point (all parameters except temperature)
Temperature Compensation

Automatic, 0 to 50°C with ß = 2%/°C (EC/TDS only)
Probe

HI 1285-5 polypropylene body, pre-amplified multiparameter probe with internal temperature sensor, 8-pin DIN connector and 1 m (3.3’) cable (included)
Battery Type / Life

9V / approximately 150 hours of continuous use
Environment

0 to 50°C (32 to 122°F); RH max 100%
Dimensions

145 x 80 x 36 mm (5.7 x 3.1 x 1.4”)
Weight

230 g (8.1 oz.)

HANNA HI 9812-5 Portable pH/EC/TDS/Temperature Meter

Darmatek

HANNA HI 9812-5 Portable pH/EC/TDS/Temperature Meter
high range EC & TDS
  • Water resistant
  • On-screen tutorial messages for calibration
  • Switch parameters at the touch of a button
  • Automatic Temperature Compensation for pH and conductivity
  • Simple calibration; Dials located on the front panel makes calibration quick and simple, even for non-technical users
  • BEPS (Battery Error Prevention System) alerts the user in the event that low battery power could adversely affect readings
  • Battery % displayed on startup
  • Fast response probe included
HI 9812-5 is a pH/EC/TDS waterproof meter designed for simplicity in taking pH, µS/cm, ppm (mg/L) and temperature measurements. The HI 9812-5 is ideal for hydroponics, greenhouses, farming and ground water applications.

Due to the built-in temperature sensor, conductivity readings are automatically compensated for temperature changes. The temperature coefficient is fixed at 2%/°C.
No probe changes are required when switching your measured parameter between pH, conductivity and TDS. These multiparameter meters reduce the number of instruments required for daily water quality analysis.

The specially engineered HI 1285-5 pH/EC/TDS and temperature probe utilizes a fiber junction and gel electrolyte which provide a fast response and reduced contamination. This combination makes the HI 1285-5 particularly designed to withstand harsh environments such as fertilizer solutions.

Order Information:

HI 9812-5 is supplied with HI 1285-5 multiparameter probe, HI 70007 pH 7.01 calibration solution sachet, HI 70032 1382 ppm (mg/L) calibration solution sachet, HI 70031 1413 µS/cm calibration solution sachet, HI 700661 electrode cleaning solution sachets (2), battery, instructions and rugged carrying case.

Specifications

Range
pH
0.0 to 14.0 pH
  EC
0 to 1990 µS/cm
  TDS
0 to 1990 ppm
  Temperature
0 to 60°C
Resolution
pH
0.1 pH
  EC
10 µS/cm
  TDS
10 mg/L
  Temperature
0°C
Accuracy
pH
±0.1 pH
  EC
±2% F.S.
  TDS
±2% F.S.
  Temperature
±1°C
TDS Conversion Factor

0.5 ppm = 1 µS/cm
Calibration

manual at 1 point (all parameters except temperature)
Temperature Compensation

Automatic, 0 to 50°C with ß = 2%/°C (EC/TDS only)
Probe

HI 1285-5 polypropylene body, pre-amplified multiparameter probe with internal temperature sensor, 8-pin DIN connector and 1 m (3.3’) cable (included)
Battery Type / Life

9V / approximately 150 hours of continuous use
Environment

0 to 50°C (32 to 122°F); RH max 100%
Dimensions

145 x 80 x 36 mm (5.7 x 3.1 x 1.4”)
Weight

230 g (8.1 oz.)

HANNA HI 2550 pH/ORP/ISE/EC/TDS/NaCl Benchtop Meter


Darmatek

HANNA HI 2550 pH/ORP/ISE/EC/TDS/NaCl Benchtop Meter
  • Up to seven measurement parameters
  • Two input channels: pH/ORP/ISE and EC/TDS/Resistivity/NaCl
  • Up to five point pH calibration with seven standard and two custom buffers
  • EC/TDS autoranging, manual  ranging and range lock
  • HOLD button to freeze readings on the display
  • Automatic Temperature Compensation (pH & EC)
  • PC interface via USB
HI 2550 is a 2 channel instrument that measures up to 7 parameters. With this single laboratory bench meter you can measure pH, ORP, ISE, conductivity (EC), TDS, NaCl percentage and temperature.

Utilizing an external temperature probe, pH readings are automatically temperature compensated. To ensure a higher level of precision, pH calibrations are up to five calibration points, chosen from the seven available memorized buffers.
This instrument can measure using ORP electrodes (pH channel input), due to it’s capability to measure mV with a resolution up to 0.1 mV and also use ISE electrodes on the mV scale (pH channel input).

EC measurements can be compensated relative to a selected reference temperature. The EC calibration mode allows you to chose from among six recognized conductivity standards and perform a single-point calibration. The most suitable EC and TDS range for your application is automatically selected. The HI 2550 also includes the ability to set and lock the range manually.

This instrument provides GLP capabilities that allows for the storage and retrieval of all data regarding pH, rel mV, EC and NaCl calibration and sample measurement as well as data regarding the maintenance and status of the electrode.
With a built-in logging function, measurements are stored in non volatile memory, and can be transferred to a PC through the USB port. Users can manually log up to 200 records and interval log up to 500 records.

Order Information:

HI 2550-01 (115V) and HI 2550-02 (230V) are supplied with HI 1131B pH electrode, HI 76310 conductivity/TDS probe, HI 7662 temperature probe, HI 76404N electrode holder, HI 70004 pH 4.01 buffer solution sachet, HI 70007 pH 7.01 buffer solution sachet, HI 7071S electrolyte solution (30 mL), 12 VDC adapter and instruction manual.

Specification

Range pH -2.0 to 16.0 pH; -2.00 to 16.00 pH; -2.000 to 16.000 pH
ISE & ORP ±999.9 mV (ISE & ORP); ±2000 mV (ISE & ORP)
EC 0.00 to 29.99 µS/cm; 30.0 to 299.9 µS/cm; 300 to 2999 µS/cm; 3.00 to 29.99 mS/cm; 30.0 to 200.0 mS/cm; up to 500.0 mS/cm actual conductivity
TDS 0.00 to 14.99 ppm; 15.0 to 149.9 ppm; 150 to 1499 ppm; 1.50 to 14.99 g/L; 15.0 to 100.0 g/L; up to 400.0 g/L actual TDS (with 0.80 factor)
NaCl 0.0 to 400.0% NaCl
Temperature –20.0 to 120.0 ºC (pH, EC range)
Resolution pH 0.1 pH; 0.01 pH; 0.001 pH
ISE & ORP 0.1 mV (±999.9 mV); 1 mV (± 2000 mV)
TDS 0.01 ppm; 0.1 ppm; 1 ppm; 0.01 g/L; 0.1 g/L
NaCl 0.1% NaCl
Temperature 0.1 ºC
Accuracy
@ 20°C
pH ±0.01 pH; ±0.002 pH
ISE & ORP ±0.2 mV (±999.9 mV); ±1 mV (±2000 mV)
EC ±1 % reading (±0.05 µS/cm or 1 digit, whichever greater)
TDS ±1% of reading (±0.03 ppm or 1 digit, whichever greater)
NaCl ±1% of reading
Temperature ±0.4 ºC (excluding probe error)
Calibration pH up to five point calibration, seven standard buffers available
(1.68, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45), and two custom buffers
EC two point calibration; one point slope calibration; six buffers available: 84.0, 1413 µS/cm; 5.00, 12.88, 80.0, 111.8 mS/cm; one point offset: 0.00 µS/cm
NaCl one point with HI 7037L standard (optional)
Probes pH HI 1131B glass body pH electrode with BNC connector and 1 m (3.3') cable (included)
EC HI 76310 platinum four ring conductivity/TDS probe with built-in temperature sensor and 1 m (3.3') cable (included)
Temperature HI 7662 temperature probe with 1 m (3.3') cable (included)
Relative mV Offset
±2000 mV
Temperature
Compensation

manual or automatic from: -20.0 to 120.0 ºC (pH range) -20.0 to 120.0 ºC (EC range) (can be disabled on conductivity range to measure actual conductivity)
Cond. Temp.
Coefficient

0.00 to 6.00 %/ºC (for EC and TDS only) default value is 1.90 %/ºC
TDS Factor
0.40 to 0.80 (default value is 0.50)
Input Impedance
1012 ohms
PC Connectivity
opto-isolated USB
Log on Demand
200 records
Log Interval Feature
500 records; 5, 10, 30 sec and 1, 2, 5, 10, 15, 30, 60, 120, 180 min stability logging (“StAb”)
Power Supply
12 VDC
Environment
0 to 50°C (32 to 122°F); RH max 95% non-condensing
Dimensions
235 x 218 x 108 mm (9.2 x 8.5 x 4.2”)
Weight