Features
- Comparator function memorizes up to seven tables.
- Dual comparator modes: Hi-Lo compares upper and lower limits, and REF-% compares a range and standard value.
- Fast response of about 100 ms (measuring pure resistance: actual response depends on material under test)
- Temperature compensation function measures temperature and calculates value relative to copper at 20°C/68°F.
- Utilizes 4-terminal method to eliminate effects of leads and contact resistance.
- Auto-ranging function.
- Dual power system: batteries or AC adapter.
- BCD, printer interface and RS-232C interface options in -01, -02 and -03 suffix versions, respectively.
Measurement ranges and Accuracy
30 mΩ to 30 k-Ohm, 7 ranges, 3500 full digits
±0.1 % rdg. ±6 dgt. (30 mΩ, 3 Ohm range)
±0.1 % rdg. ±4 dgt. (300 mΩ, 30 Ohm to 30 k-Ohm range) - See more at: http://www.tridinamika.com/electronic-measuring-detail/hioki-3540-01/02/03-milliohm-hitester#sthash.2w7BfoUr.dpuf
±0.1 % rdg. ±6 dgt. (30 mΩ, 3 Ohm range)
±0.1 % rdg. ±4 dgt. (300 mΩ, 30 Ohm to 30 k-Ohm range) - See more at: http://www.tridinamika.com/electronic-measuring-detail/hioki-3540-01/02/03-milliohm-hitester#sthash.2w7BfoUr.dpuf
Measurement ranges and Accuracy
30 mΩ to 30 k-Ohm, 7 ranges, 3500 full digits
±0.1 % rdg. ±6 dgt. (30 mΩ, 3 Ohm range)
±0.1 % rdg. ±4 dgt. (300 mΩ, 30 Ohm to 30 k-Ohm range) - See more at: http://www.tridinamika.com/electronic-measuring-detail/hioki-3540-01/02/03-milliohm-hitester#sthash.2w7BfoUr.dpuf
Measurement ranges and Accuracy
:±0.1 % rdg. ±6 dgt. (30 mΩ, 3 Ohm range)
±0.1 % rdg. ±4 dgt. (300 mΩ, 30 Ohm to 30 k-Ohm range) - See more at: http://www.tridinamika.com/electronic-measuring-detail/hioki-3540-01/02/03-milliohm-hitester#sthash.2w7BfoUr.dpuf
30mΩ to 30kΩ, 7 Ranges, 3500 Full Digits
±0.1 % rdg. ±6 dgt. (30mΩ, 3Ω Range)
±0.1 % rdg. ±4 dgt. (300mΩ, 30Ω to 30kΩ Range)
Measurement current :
100 mA (30mΩ, 300 mΩ Range)
Max. applied measurement: 3.5 mV DC (30 mΩ Range) to 350 mV DC (30kΩ Range)
Sampling Speed :
Fast: 16 times/second
Slow: 4 times/second
Response Time :
Fast: 100ms
Slow: 300ms
Display :
3500 full digits, Liquid Crystal Display
Measurement method : Four-terminal measurement
Open-circuit terminal voltage :
4.0 V Max. (30 mΩ to 30 kΩ all ranges)
Digital input/output (-01, -02 and -03 Ver. only) :
TTL output BCD, or other inputs/outputs for external control
Comparator functions : Setting:Upper and lower limit, or reference value and % for resistance, Up to 7 tables
Output:3 levels (Hi, In, Lo), Open-collector, LED display, beep sound
Power supply : LR6 (AA) or R6P (AA) × 6, or (OPTIONAL) 9445-02, -03 AC ADAPTER (9 V, 1.4 A)