HANNA HI 98150 Portable pH/ORP Meter with SMART Electrode
- pH Calibration Check™ Electrode Diagnostics
- Waterproof
- Up to five point pH calibration
- Up to seven automatically recognized pH buffers and up to two custom pH buffers
- Log on demand (500 samples)
- GLP features
- Auto endpoint
- User-selectable “calibration time out”
- Backlit, multi-level LCD display
- Tutorial messages on LCD
- Opto-isolated USB
- BEPS and % battery life on display; Battery Error Prevention System alerts the user in the event that low battery power could adversely affect readings
SMART electrodes incorporate a chip, which stores the calibration data performed with a specific instrument. When the “SMART” electrode is attached to the meter again, it is automatically recognized. This allows the operator to optimize time and efficiency with unsurpassed safety. It also avoids erroneous measurements taken in the event the pH electrode is substituted after calibration.
Order Information:
HI 98150 is supplied with HI 1618D pH electrode, HI 70004 pH 4.01 buffer solution sachet, HI 70007 pH 7.01 buffer solution sachet, HI 70000 electrode cleaning solution sachet, HI 700661 electrode cleaning solution sachet, batteries, instructions and hard carrying case.
Specification
Range |
pH |
-4.0 to 20.0 pH; -4.00 to 20.00 pH |
ORP |
±699.9; ±2000 mV |
|
Temperature |
-20.0 to 120.0°C (-4.0 to 248.0°F) |
|
Resolution |
pH |
0.1 pH; 0.01 pH |
ORP |
0.1 mV (±699.9 mV); 1 mV (outside) |
|
Temperature |
0.1°C (0.1°F) |
|
Accuracy @ 20°C |
pH |
±0.1 pH; ±0.01 pH |
ORP |
±0.2 mV (±699.9 mV); ±1 mV (±2000 mV) |
|
Temperature |
±0.2°C (±0.4°F) excluding probe error |
|
Calibration |
pH |
up to five point calibration with seven standard buffers available
(1.68, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45), and two custom buffers |
Offset |
±1 pH |
|
Slope |
from 80 to 110% |
|
Temperature Compensation |
|
manual or automatic from –20.0 to 120.0 ºC (–4.0 to 248.0 ºF) |
pH Electrode |
|
HI 1618D PEI body, pre-amplified pH electrode with “SMART”
technology, internal temperature sensor, DIN connector and 1 m (3.3’)
cable (included) |
Logging |
|
log on demand, 500 samples |
PC Connectivity |
|
opto-isolated USB with optional HI 92000 software |
Input Impedance |
|
1012 Ohms |
Battery Type / Life |
|
1.5V AAA (3) / approximately 200 hours of continuous use without
backlight (50 hours with backlight). User selectable auto-off (5, 10,
20, 60 minutes or can be disabled) |
Environment |
|
0 to 50°C (32 to 122°F); RH max 100% |
Dimensions |
|
185 x 72 x 36 mm (7.3 x 2.8 x 1.4”) |
Weight |
|
300 g (10.6 oz.) |
0 komentar:
Posting Komentar