HANNA HI 98150 Portable pH/ORP Meter with SMART Electrode

Darmatek

HANNA HI 98150 Portable pH/ORP Meter with SMART Electrode
  • pH Calibration Check™ Electrode Diagnostics
  • Waterproof
  • Up to five point pH calibration
  • Up to seven automatically recognized pH buffers and up to two custom pH buffers
  • Log on demand (500 samples)
  • GLP features
  • Auto endpoint
  • User-selectable “calibration time out”
  • Backlit, multi-level LCD display
  • Tutorial messages on LCD
  • Opto-isolated USB
  • BEPS and % battery life on display; Battery Error Prevention System alerts the user in the event that low battery power could adversely affect readings
HI 98150 is a portable Calibration Check™ pH meter that utilizes HANNA’s SMART electrode technology.
SMART electrodes incorporate a chip, which stores the calibration data performed with a specific instrument. When the “SMART” electrode is attached to the meter again, it is automatically recognized. This allows the operator to optimize time and efficiency with unsurpassed safety. It also avoids erroneous measurements taken in the event the pH electrode is substituted after calibration.

Order Information:
HI 98150 is supplied with HI 1618D pH electrode, HI 70004 pH 4.01 buffer solution sachet, HI 70007 pH 7.01 buffer solution sachet, HI 70000 electrode cleaning solution sachet, HI 700661 electrode cleaning solution sachet, batteries, instructions and hard carrying case.

Specification

Range
pH
-4.0 to 20.0 pH; -4.00 to 20.00 pH
  ORP
±699.9; ±2000 mV
  Temperature
-20.0 to 120.0°C (-4.0 to 248.0°F)
Resolution
pH
0.1 pH; 0.01 pH
  ORP
0.1 mV (±699.9 mV); 1 mV (outside)
  Temperature
0.1°C (0.1°F)
Accuracy @ 20°C
pH
±0.1 pH; ±0.01 pH
  ORP
±0.2 mV (±699.9 mV); ±1 mV (±2000 mV)
  Temperature
±0.2°C (±0.4°F) excluding probe error
Calibration
pH
up to five point calibration with seven standard buffers available (1.68, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45), and two custom buffers
  Offset
±1 pH
  Slope
from 80 to 110%
Temperature
Compensation


manual or automatic from –20.0 to 120.0 ºC (–4.0 to 248.0 ºF)
pH Electrode

HI 1618D PEI body, pre-amplified pH electrode with “SMART” technology, internal temperature sensor, DIN connector and 1 m (3.3’) cable (included)
Logging

log on demand, 500 samples
PC Connectivity

opto-isolated USB with optional HI 92000 software
Input Impedance

1012 Ohms
Battery Type / Life

1.5V AAA (3) / approximately 200 hours of continuous use without backlight (50 hours with backlight). User selectable auto-off (5, 10, 20, 60 minutes or can be disabled)
Environment

0 to 50°C (32 to 122°F); RH max 100%
Dimensions

185 x 72 x 36 mm (7.3 x 2.8 x 1.4”)
Weight

300 g (10.6 oz.)