Minggu, 17 April 2016

Kyoritsu 3128 12kV H. V. Insulation Tester


Kyoritsu 3128 12kV H. V. Insulation Tester
  • Microprocessor controlled high voltage insulation resistance tester with Diagnostic functions.
  • Suitable for analyzing the insulation characteristics of cables, transformers, motors, generators, high-pressure switches, insulators, wiring installations, etc.
  • 6 ranges: 500V, 1000V, 2500V, 5000V, 10000V, 12000V
  • Fine adjustment of voltage setting at each range is also possible
  • Graphic representation of the insulation resistance and leakage current versus time on large display with bar graph and backlight.
  • Can be operated from built-in rechargeable battery or from AC line.
  • Automatic discharge after test with monitoring of the discharge voltage.
  • Internal memory can store about 43,000 data (max).
  • Robust design for field use with IP64 rating (with lid closed).
Specification
Insulation resistance (*) : Accuracy is not guaranteed with setting of 250V or less
Rated voltage 500V 1000V 2500V 5000V 10000V 12000V
Max. measurement value 500GΩ 1TΩ 2.5TΩ 5TΩ 35TΩ




Accuracy
400k~50GΩ
±5%rdg±3dgt
(*)
800k~100GΩ
±5%rdg±3dgt
2M~250GΩ
±5%rdg±3dgt
4M~500GΩ
±5%rdg±3dgt
8M~1TΩ
±5%rdg±3dgt
50G~500GΩ
±20%rdg
(*)
100G~1TΩ
±20%rdg
250G~2.5TΩ
±20%rdg
500G~5TΩ
±20%rdg
1T~10TΩ
±20%rdg
10T~35TΩ
Values are displayed, but accuracy isn't guaranteed
Short circuit current Max 5.0mA
Load resistor to output rated voltage 0.5MΩ or more 1MΩ or more 2.5MΩ or more 5MΩ or more 20MΩ or more 24MΩ or more
Output Voltage
Rated Voltage 500V 1000V 2500V 5000V 10000V 12000V
Monitor Accuracy ±10%±20V ±10%±20V ±10%±20V ±10%±20V ±10%±20V ±10%±20V
Output Accuracy 0~+20% 0~+10% 0~+10% 0~+10% -5~+5% -5~+5%
Selectable range 50~600V
(in steps of 5V)
610~1200V
(in steps of 10V)
1225~3000V
(in steps of 25V)
3050~6000V
(in steps of 50V)
6100~10000V (in steps of 100V) 10100~
12000V
(in steps of 100V)
Voltage measurement
Measuring range DCV : ±30~±600V, ACV : 30~600V (50/60Hz)
Accuracy ±2%rdg±3dgt
Current measurement
Measuring range 0.00nA~2.40mA (Depending on the insulation resistance)
Capacitance measurement
Measuring range 5.0nF~50.0µF 5.0nF~1.0µF
(display range: 5.0nF~50.0µF)
Accuracy ±5%rdg±5dgt
General
Withstand voltage AC8770V : between line terminal and enclosure / 5sec (50/60Hz)
AC6880V : between the measuring terminal and enclosure / 5sec (50/60Hz)
AC2330V : between the power connector and enclosure / 5sec (50/60Hz)
Operating temperature & humidity range -10°C ~ 50°C /Relative humidity 85% or less
(when operating with an external power supply, no condensation)
0°C ~ 40°C /Relative humidity 85% or less
(when operating with battery, no condensation)
Storage temperature & humidity range
-20ºC~60ºC / Relative humidity 75% or less (no condensation)
Applicable standards IEC 61010-1 CAT.IV 600V Pollution degree 2,
IEC 61010-031, IEC 61326, IEC 60529 (IP64): with the lid closed.



Power source
Rechargeable Lead storage battery (12V)*
Charging time: approx. 8 hours / AC Power supply (100V~240V, 50/60Hz)

* Continuous measuring time:
approx. 4 hours
(a load of 100MΩ at the Insulation resistance 12000V Range.)
approx. 2 hours
(Max current consumption in range while rated currents are maintained.)
Dimension 330 (L) × 410 (W) × 180 (D) mm *Instrument and Hard case
Weight 9kg approx. (including battery) *Instrument and Hard case
Accessories 7170 (Main cord), 7224A (Earth cord), 7225A (Guard cord), 7226A (Line probe), 7227A (Line probe with alligator clip), 8029 (Extension prod), 8212-USB-W (USB adaptor with KEW Windows(Software)), Instruction manual, Calibration Certificate
Optional 7254 (Longer Line probe with alligator clip)

0 komentar: