Minggu, 17 April 2016

SEW 6213A-IN Digital H.V. Insulation Tester


SEW 6213A-IN Digital H.V. Insulation Tester

FEATURES
  • Microprocessor-controlled.
  • 2 ×16 characters; large intelligent LCD module.
  • 20 Insulation test voltages 500V, 1kV, 1.5kV, 2kV, 2.5kV, 3kV, 3.5kV, 4kV, 4.5kV, 5kV, 5.5kV, 6kV, 6.5kV, 7kV, 7.5kV, 8kV, 8.5kV, 9kV, 9.5kV, 10kV.
  • Ener-Save™ feature to extend battery life.
  • Bar-graph indicates test voltage; rise and decay can be observed.
  • Visual and audible warning if external voltage is present (>500Vac or Vdc.)
  • Insulation resistance; auto-ranging on all ranges.
  • Overload protection.
  • Low battery indicator (read time battery voltage measurement).
  • Measures insulation time duration of the test.
  • Low battery consumption.
  • Better than 10% accuracy on all ranges.
  • Auto-off.
  • Calculates Dielectric Absorption Ratio (DAR) automatically.
  • Calculates Polarization Index (PI) automatically.
SPECIFICATIONS
Test voltage  From 500Vdc to 10kVdc
 Adjustable in 500V steps
 Preset buttons  1kV / 2.5kV / 5kV / 10kV
 Measuring ranges  800kΩ-500GΩ (auto-ranging)
 Accuracy  ± 5% ± 2 digits
 Output power limit  1W
 Voltage regulation  Selected voltage +20%-5% of
 nominal value unless current limited.
 Meaning that if output current is
 too high, the voltage will be lowered 
 automatically.
 Dimensions  330(L) × 260(W) × 160(D)mm
 Weight
 (battery included)
  Approx. 3600g
 Power source  1.5V “C”× 8 Alkaline batteries
 Safety standard  EN 61010-1 CAT III  300V
 EN 61326-1